Quantification of the Multiplexing Effects in Multi-Pinhole Small Animal SPECT: A Simulation Study
Publication in refereed journal

香港中文大學研究人員
作者已離職


引用次數
替代計量分析
.

其它資訊
摘要Our goal is to study the trade-off between image degradation and improved detection efficiency and resolution from allowing multiplexing in multi-pinhole (MPH) SPECT, and to determine The optimal pinhole number for; MPH design. We used an analytical 3D MPH projector and two digitized phantoms: the mouse whole body (MOBY) phantom and a hot sphere phantom to generate noise-free and noisy projections, simulating pinhole collimators fitted with pre-studied pinhole patterns. We performed three schemes to achieve different degrees of multiplexing: 1. Fixed magnification and detection efficiency; 2. Fixed detection efficiency and changed magnification; 3. Fixed magnification and changed detection efficiency. We generated various noisy data sets by simulating Poisson noise using differently scaled noise-free projections and obtained 20 noise realizations for each setting. All datasets; were reconstructed using 3D MPH ML-EM reconstruction method. We analyzed the quantitative accuracy by the normalized-mean-square-error. We evaluated the image contrast for the hot sphere phantom simulation, and also the image noise by the average normalized-standard-deviation of certain pixels for different degrees of multiplexing. Generally no apparent artifacts were observed in the reconstructed images, illustrating the effectiveness of reconstructions. Bias increased for increased degree of multiplexing. Contrast was not significantly affected by multiplexing in the specific simulation scheme (1). Scheme (2) showed that excessive multiplexing to improve image resolution would not improve the overall trade-off of bias and noise compared to no multiplexing. However, scheme (3) showed that when comparing to no multiplexing, the trade-off improved initially with increased multiplexing by allowing more number of pinholes to improve detection efficiency. The trade-off reached a maximum and decreased with further multiplexing due to image degradation from increased bias. The optimal pinhole number was 7 for a compact camera with size of 12 cm x 12 cm and 9 for a standard. gamma camera with size of 40 cm x 40 cm in this scheme. We conclude that the gains in improved detection efficiency and resolution by increased multiplexing are offset by increased image degradations. All the aforementioned factors must be considered in the optimum MPH collimator design for small animal SPECT imaging.
著者Mok GSP, Wang YC, Tsui BMW
期刊名稱IEEE Transactions on Nuclear Science
出版年份2009
月份10
日期1
卷號56
期次5
出版社Institute of Electrical and Electronics Engineers (IEEE)
頁次2636 - 2643
國際標準期刊號0018-9499
語言英式英語
關鍵詞Multi-pinhole collimator; multiplexing; small animal imaging; SPECT
Web of Science 學科類別Engineering; Engineering, Electrical & Electronic; ENGINEERING, ELECTRICAL & ELECTRONIC; Nuclear Science & Technology; NUCLEAR SCIENCE & TECHNOLOGY

上次更新時間 2020-17-11 於 23:37