A Novel Parameter from PPG Dicrotic Notch for Estimation of Systolic Blood Pressure Using Pulse Transit Time
Refereed conference paper presented and published in conference proceedings


Full Text

Times Cited
Web of Science17WOS source URL (as at 17/10/2020) Click here for the latest count

Other information
AbstractThis paper proposes a new calibration parameter from the dicrotic notch of photoplethysmography (PPG) waveform for systolic blood pressure (SBP) estimation using pulse transit time (PTT). An experiment including exercise was conducted on twelve subjects. The results show that estimation with PTT and a new parameter, i.e. the Relative Amplitude of Secondary peak (RAS), can predict SBP within 1.7+/-6.8 mmHg of the reference for all the subjects before and after exercise (including within 5 minutes and 20-40 minutes after exercise). The underestimated bias after exercise if PTT was used alone can be largely reduced when RAS is introduced. The preliminary analysis indicates that RAS is a promising parameter for calibrating the PTT-based approach for cuffless BP estimation.
All Author(s) ListGu WB, Poon CCY, Zhang YT
Name of Conference5th International Summer School and Symposium on Medical Devices and Biosensors
Start Date of Conference01/06/2008
End Date of Conference03/06/2008
Place of ConferenceHong Kong
Country/Region of ConferenceChina
Detailed descriptionorganized by The Chinese University of Hong Kong,
Year2008
Month1
Day1
PublisherIEEE
Pages292 - 294
ISBN978-1-4244-2252-4
LanguagesEnglish-United Kingdom
Web of Science Subject CategoriesEngineering; Engineering, Biomedical; Engineering, Electrical & Electronic