Time caliper method with enhanced accuracy using an arbitrary waveform generator and time demultiplexer with fine clock offset
Publication in refereed journal

香港中文大學研究人員

引用次數
替代計量分析
.

其它資訊
摘要An improved time caliper method has been proposed for measuring a small relative delay of two signals transmitted on different channels. This is achieved by using a new method to generate two periodic pulse trains that have much smaller period difference, over 1000 times smaller than that in the prior scheme. The two pulse trains, with a period difference of 0.1 ps, are generated by an arbitrary waveform generator and a time demultiplexer, with a fine clock offset. The average and maximum measurement errors obtained are 0.68 and 1.67 ps, respectively. (C) 2014 Optical Society of America
著者Luk HT, Chen LK
期刊名稱Optics Letters
出版年份2014
月份10
日期15
卷號39
期次20
出版社Optical Society of America
頁次5965 - 5968
國際標準期刊號0146-9592
電子國際標準期刊號1539-4794
語言英式英語
Web of Science 學科類別Optics

上次更新時間 2020-23-11 於 00:06