A new robotically assisted system for total knee arthroplasty: A sheep model study
Publication in refereed journal

替代計量分析
.

其它資訊
摘要Background: We investigated the accuracy and safety of a new HURWA robotic-assisted total knee arthroplasty (TKA) system in a sheep model. Methods: Ten male small-tailed Han sheep were used in this study. Sheep were imaged by computed tomography scan before and after bone resection and the cutting errors between actual bone preparation and preoperative planning of the femur and tibia in three dimensions were measured. Results: The overall accuracies after surgery compared with that from preoperative surgical planning of the left and right femurs were 1.93 ± 1.02° and 1.93 ± 1.23°, respectively. Additionally, similarly high overall accuracies for the left and right tibia of 1.26 ± 1.04 and 1.68 ± 0.92°, respectively, were obtained. The gap distances of the distal cut, anterior chamfer, anterior cut, posterior chamfer and posterior cut on the medial side were 0.47 ± 0.35 mm, 0.41 ± 0.37 mm, 0.12 ± 0.26 mm, 0.41 ± 0.44 mm and 0.12 ± 0.23 mm, respectively. No intraoperative complications, such as intraoperative fracture, massive bleeding or death, occurred. Conclusion: This new HURWA robotic-assisted TKA system is an accurate and safe tool for TKA surgery based on the sheep model.
出版社接受日期09.04.2021
著者Chen X, Li Z, Zhang X, Yan J, Ding L, Song Y, Huo Y, Chan MTV, Wu WKK, Lin J
期刊名稱International Journal of Medical Robotics and Computer Assisted Surgery
出版年份2021
月份8
卷號17
期次4
出版社Wiley
出版地England
文章號碼e2264
國際標準期刊號1478-5951
電子國際標準期刊號1478-596X
語言英式英語

上次更新時間 2024-21-08 於 00:42